Product name:Liquid Based Cytology Specimen Production System

Introduction:

  • Patented innovation

  • User-friendly design

  • One-step specimen production

  • TBS standardized reporting system

  • Versatile configuration

  • Applicable to various cytology production (such as cervical exfoliated cells, sputum, hydrothorax, cerebrospinal fluid, etc.)

Inquiry


Operating Procedure

Sampling

1. Insert the brush into the cervix and ensure the brush head close to the cervix;

2. Slowly rotate over 5 turns axially;

Cell Collection

1. Remove the brush head and put it into the cell preservation solution with the cap tightened;

2. Shake more than 15 times by hand or put it on oscillator to oscillate for 10 minutes;

Cell Production

1. Draw 0.5-3ml mixed solution with pipette onto the production holder which should be placed symmetrically;

2. Produce by 1000 rpm or 1600 rpm/min centrifugal speed for 4-5 minutes;

Dyeing

1. Remove the glass sides with cell smear out of the solution holder;

2. Put the slider into 95% alcohol and fix it for 10-15mins to stain. 